Cannot operate MS instrument fitted with UniSpray source enclosure - WKB45025
Article number: 45025
SYMPTOMS
- Tune page does not recognize that the UniSpray source is fitted
- Collision gas flow control slider is not present on the tune page
- Control over the collision gas flow rate during acquisition is compromised
- MS instrument fitted with API source enclosure and ESI probe can be controlled as expected
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ Absolute
- Synapt XS
- UniSpray
- MassLynx 4.2, SCN 986
CAUSE
Either not officially supported or bugs in the software, including lack of automatic recognition of the UniSpray source
FIX or WORKAROUND
Resolved for Xevo TQ-S in SCN 997 and subsequent software - see article What is the latest version of MassLynx required for a Xevo TQ-S? - WKB5508
Workaround for earlier versions and for Synapt XS:
- Fit the UniSpray source enclosure.
- Manually select ESI+ or ESI- in the source drop-down list of the tool bar menu on the MS tune page.
- Use the "capillary" voltage slider to control the impactor voltage.
- Note the Experimental record with record the ion mode as ESI and the impactor voltage as capillary voltage