Cannot operate MS instrument fitted with UniSpray source enclosure - WKB45025

Article number: 45025

SYMPTOMS

  • Tune page does not recognize that the UniSpray source is fitted
  • Collision gas flow control slider is not present on the tune page
  • Control over the collision gas flow rate during acquisition is compromised
  • MS instrument fitted with API source enclosure and ESI probe can be controlled as expected

ENVIRONMENT

  • Xevo TQ-S
  • Synapt XS
  • UniSpray
  • MassLynx 4.2, SCN 986

CAUSE

Either not officially supported or bugs in the software, including lack of automatic recognition of the UniSpray source

FIX or WORKAROUND

Resolved for Xevo TQ-S in  SCN 997 and subsequent software - see article What is the latest version of MassLynx required for a Xevo TQ-S? - WKB5508

Workaround for earlier versions and for Synapt XS:

  1. Fit the UniSpray source enclosure.
  2. Manually select ESI+ or ESI- in the source drop-down list of the tool bar menu on the MS tune page.
  3. Use the "capillary" voltage slider to control the impactor voltage.
  4. Note the Experimental record with record the ion mode as ESI and hte impactor voltage as capillary voltage 

ADDITIONAL INFORMATION

id45025, Electrospray, MLYNX, MLYNXV41, SCN986, SCN997, SUPMM, SYNAPTXS, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD