Cannot operate MS instrument fitted with UniSpray source enclosure - WKB45025
SYMPTOMS
- Tune page does not recognize that the UniSpray source is fitted
- Collision gas flow control slider is not present on the tune page
- Control over the collision gas flow rate during acquisition is compromised
- MS instrument fitted with API source enclosure and ESI probe can be controlled as expected
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ Absolute
- Synapt XS
- UniSpray
- MassLynx 4.2, SCN 986
- waters_connect
CAUSE
Either not officially supported or bugs in the software, including lack of automatic recognition of the UniSpray source
FIX or WORKAROUND
Resolved for Xevo TQ-S in SCN 997 and subsequent software - see article What is the latest version of MassLynx required for a Xevo TQ-S? - WKB5508
Workaround for earlier versions and for Synapt XS:
- Fit the UniSpray source enclosure.
- Manually select ESI+ or ESI- in the source drop-down list of the tool bar menu on the MS tune page.
- Use the "capillary" voltage slider to control the impactor voltage.
- Note the Experimental record with record the ion mode as ESI and hte impactor voltage as capillary voltage
ADDITIONAL INFORMATION
CRI-8750 has been created for this issue on Waters Connect including W_C 4.0. Driver 1.7 and below
The workaround for Masslynx also works in W_C but additionally if the LND (workstation LND reset or network reset) has to be reset it will trigger the issue again and require the workaround to be completed again