Persistent error message "Stopped - Select purge to recover" for Vion IMS QTof - WKB4476
Article number: 4476
SYMPTOMS
- "Stopped - Select purge to recover" error message appears in fluidics progress bar
- Health status shows leak detected
ENVIRONMENT
- Vion IMS QTof
- SYNAPT G2-Si
- Xevo G2-S
- Xevo G2-XS
CAUSE
Leak sensor of the MS fluidics detected a leak.
FIX or WORKAROUND
- Inspect for leaks around the leak tray.
- Tighten loose fittings and repair any leaks.
- Dry out the tray and the leak sensor.
ADDITIONAL INFORMATION
The instrument can be purged when the leak detected health status has been resolved. If there is no leak, the leak sensor cable might not be properly plugged in, or it might be broken.