Skip to main content
Waters

Persistent error message "Stopped - Select purge to recover" for Vion IMS QTof - WKB4476

Article number: 4476

SYMPTOMS

  • "Stopped - Select purge to recover" error message appears in fluidics progress bar
  • Health status shows leak detected

ENVIRONMENT

  • Vion IMS QTof
  • SYNAPT G2-Si
  • Xevo G2-S
  • Xevo G2-XS

CAUSE

Leak sensor of the MS fluidics detected a leak.

FIX or WORKAROUND

  1. Inspect for leaks around the leak tray. 
  2. Tighten loose fittings and repair any leaks.
  3. Dry out the tray and the leak sensor.

ADDITIONAL INFORMATION

The instrument can be purged when the leak detected health status has been resolved. If there is no leak, the leak sensor cable might not be properly plugged in, or it might be broken.

Not able to find a solution? Click here to request help.