Skip to main content
Waters

Calculation of USP Signal-to-Noise Ratio - Tip201

Article number: 118410

OBJECTIVE or GOAL

Get Empowered: Empower User Question

Tip #201: Calculation of USP Signal-to-Noise Ratio

The previous tip (Tip #200) discussed a question about identifying peaks in Empower.  

This tip discusses the calculation of USP signal-to-noise ratio.

ENVIRONMENT

  • Empower

PROCEDURE

  1. The result displayed does not have the USP signal-to-noise (s/n) nor the USP Noise calculated.
    118410_1.png
     
  2. The ‘Suitability’ tab in the Processing Method shows the correct selection for calculating USP signal-to-noise (s/n) and the use of a blank injection to calculate noise.
    118410_2.png
     
  3. Working backwards, ‘Alter Sample Set’ displays the ‘Sample Matrix’ field.  The problem is the selection for the ‘Blank’ is different from the rest of the samples.
    118410_3.png
     
  4. Modify the ‘Sample Matrix’ to the same as the blank.
    118410_4.png
     
  5. Reprocess the Sample Set and now the result for the same sample shows the values are calculated.
    118410_5.png
     
  6. Search on ‘Verifying System Performance’ in the online help and you will see a Tip regarding the Sample Matrix.
    118410_6.png

ADDITIONAL INFORMATION

Final Note: Refer to Tip #28 and Tip #30 for more detail on calculating signal-to-noise (s/n) ratio

Not able to find a solution? Click here to request help.