The 7697 control panel is blank on the Run Samples screen through Citrix and Empower - WKB7872

Article number: 7872

SYMPTOMS

  • The 7697A control panel is blank in the Run Samples screen when accessing system through Citrix
  • The 7697A control panel is normal in the Run Samples screen when accessing system directly through the LAC/E

ENVIRONMENT

CAUSE

Feature release or service release on the Citrix Application Server is not at the minimum for Headspace Control Option 3.0. 

 

HeadspaceControlOption3.0RN.jpg

 

FIX or WORKAROUND

  1. Upgrade to Empower 3 Service Release 2 Hotfix 1.

ADDITIONAL INFORMATION

See related article WKB24744.

See article WKB62833 on split injection technique.

See article WKB62831 on splitless injection technique.

See article WKB62933 on PTV injection technique.

 