Cannot monitor the GC method in Empower - WKB24744

Article number: 24744

SYMPTOMS

  • Cannot monitor the baseline, but no errors observed
  • Can make an injection without issue 

ENVIRONMENT

  • Empower 3 SR2 HF1 Enterprise
  • Citrix
  • Agilent GC 7890A
  • Agilent Headspace G1888
  • Agilent GC ICS v2.5
  • Waters Headspace Control Option v2.0 and v3.0

CAUSE

The Citrix server has Headspace control option v3.0, but the LAC/E has Headspace control option v2.0.

FIX or WORKAROUND

Install the matching ICS.

ADDITIONAL INFORMATION

See related article WKB7872

See article WKB62833 on Split injection technique

See article WKB62831 on Splitless injection technique

See article WKB62933 on PTV injection technique

id24744, 7890 GC, 7890 GCDET, 7890BGC, HEADSPACE