The shutdown method on the General tab of the instrument method editor appears fully functional for some Agilent LCs but is missing for some using Empower 3 - WKB68563
Article number: 68563
SYMPTOMS
- The shutdown method on the General tab of the instrument method editor appears fully functional for some Agilent LCs but is missing for some using Empower 3
- There is no correlation to firmware levels
- Systems were created at different times using the Agilent PreConfiguration utility
ENVIRONMENT
- Empower 3 Feature Release 2
- Agilent Instrument Control Framework (ICF) Support Version 3.0
- Agilent 1290 Infinity II LC
- Agilent PreConfiguration utility
CAUSE
- The Agilent PreConfiguration utility was used to create the systems, but usage was inconsistent
- Some systems were created using the Tools menu in Configuration Manager; others were created using the Windows folder (C:\empower\instruments\Agilent)
FIX or WORKAROUND
- Delete and reconfigure the instrument using the Agilent PreConfiguration utility.
- Use the Agilent PreConfiguration utility from the Tools in Configuration Manager instea of from the Windows folder, and be consistent for all systems.
ADDITIONAL INFORMATION
- Since Agilent ICF 2.2, it is recommended that you use the Tools drop-down menu. Only use the Windows drop-down menu with Agilent ICF 2.1 A.02.03 DU1 HF2.
- Article WKB13860 - How do you determine which process was used for the Agilent PreConfiguration utility?