Error: The report method cannot be used to sign off results in Empower - WKB50273

Article number: 50273

SYMPTOMS

  • Trying to perform e-signature with the appropriate privilege brings the error "The report method [name of method] cannot be used to sign off results" (see screen shot below)

sign off error.PNG

ENVIRONMENT

  • Empower 3 
  • Self-directed qualification report "Noise RM RS" and "Flow Rate Linearity RM RS" for UPLC QSM-FTN-PDA project

CAUSE

The report method type is not a sign-off report method.

FIX or WORKAROUND

  1. Open the report method.
  2. Right-click on a blank section in the report and click Method Properties.
  3. Select the "Allow this method to be used in Sign Off" check box to convert the report to a sign-off report. 
  4. Save the report method.

sign off report.PNG

ADDITIONAL INFORMATION

WKB27317

This defect was reported in January 2016 as PCS #24633. It's also recorded under PCS #48765 for Qualification reports.