The check box in Empower report method properties to "allow method to be used in sign off" is unavailable (dimmed) - WKB27317

Article number: 27317

SYMPTOMS

  • The check box in report method properties to "allow method to be used in sign off" is unavailable (dimmed)
  • User cannot enable method to be used as sign off

image.png

ENVIRONMENT

  • Empower 

CAUSE

User does not have the privilege "specify report methods for sign off".

FIX or WORKAROUND

Assign user the privilege "specify report methods for sign off", or allow an administrator with this privilege to allow the report method to be used for sign-off.

ADDITIONAL INFORMATION

 