Loss of communication between 2695 Separations Module and Empower - WKB239722

Article number: 239722

SYMPTOMS

  • Loss of communication between Alliance system and Empower

ENVIRONMENT

  • Empower 3
  • 2695 Separations Module

CAUSE

The Instrument LAN IP address is missing.

FIX or WORKAROUND

Reconfigure the Instrument LAN settings by following the article "How To Configure the Waters DHCP Server to control the Ethernet instruments".

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.