SQD instrument software communication failure when opening - WKB194906
Article number: 194906
SYMPTOMS
- SQD instrument software communication failure when opening
- Ping of the instrument OK
- Connection to telnet epc OK
- Not possible to control SQD
- When opening Console, no serial number, calibration state, all is missing
ENVIRONMENT
- Empower
- Windows 20H2
- SQD
- Empower 3 FR4 SR3 HF1
- SQD driver: 1.44
- Waters driver pack: 2019 release 1
- Firewall is completely disabled; the exclusions are not set
CAUSE
The issue started when updating to Windows 10 build 20H2.
Reverting to Windows version used previously (build 1909) resolved the issue.
FIX or WORKAROUND
- Revert to a known working Windows 10 version.
- Recommendaton: Install supported Windows 10 version (currently v1607 LTSB).
ADDITIONAL INFORMATION
- All steps in WKB99782 attempted.
- Removing and adding instruments from DHCP one by one did not resolve the issue.
id194906, A-SQ, comms, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SQD, SUP