Cannot alter a running sample set in Empower - WKB14505

Article number: 14505

SYMPTOMS

  • In the Empower run sample screen, the "alter running samples" feature under Edit is unavailable. 
  • Necessary privilege to perform alter running samples function is available for the User Type.

ENVIRONMENT

  • Empower 3 Feature Release 3 (FR3)

CAUSE

Corrupted Sample Set method.

FIX or WORKAROUND

  1. Create a new Sample Set method.

ADDITIONAL INFORMATION