Article number: 14489

SYMPTOMS

  • Run stops randomly
  • The Instrument Server exits unexpectedly.
  • Instrument Server crash in Application log
  • Faulting module can be reported as MFC71U.dll  or ntdll.dll  or mil0.dll or mil1.dll

MFC71U.dll error.PNG

ENVIRONMENT

  • Empower 3
  • Acquistion
  • Instrument Server Crash

CAUSE

Defect reported as  PCS 55077 ,53915, 55005, 55040, 55536, 55595, and 55657.

FIX or WORKAROUND

Fixed in Empower 3 FR3 Hotfix 1 under PCS 55077

MFC71U.dll message with Empower 3 instrument server : Waters

ADDITIONAL INFORMATION

If  faulting module is reported  as OraClientxx.dll. This is due to PCS 55040.

See KCS  for further details

 The application Instrumentserver.exe is faulting (specifically module OraClientxx.dll) causing sample sets to crash in Empower - WKB8068 - Waters