Audit trail entries were not recorded in Empower 3 during a run - WKB14480

Article number: 14480

SYMPTOMS

  • During a run, the LAC/E32 is buffering and no audit records are captured
  • When viewing Audit trail for Sample Set error "No rows are found You may want to change conditions specified in current view filter"

ENVIRONMENT

  • Empower 3 Enterprise
  • Feature Release 2
  • Feature Release 3
  • Feature Release 4
  • Feature Release 5
  • Empower 3.6.0
  • Empower 3.6.1
  • Empower 3.7
  • Empower 3.7.1
  • Empower 3.8

CAUSE

The LAC/E32 does not hold audit records like it does data files.

FIX or WORKAROUND

This issue is documented under CRI-246 (PCS 54033) and is not fixed as of Empower 3.8