The PDA/BSM control panel is missing from the Empower Run Samples window - WKB122465

Article number: 122465

SYMPTOMS

  • The PDA control panel is missing from the Empower Run Samples window
  • IP addresses populate with correct information in DHCP when instrument is removed 
  • Stand-alone ACQUITY Console shows "Communication Failure"

ENVIRONMENT

  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC BSM

CAUSE

ICS corrupted

FIX or WORKAROUND

Use the Repair function within Deployment Manager (see WKB206867).                                                                                                                                                                                     clipboard_e03ba98bc59e40e1fbd79e1e4636668fb.png