Communications loss with SQD and Empower when using remote desktop and touchscreen PC with Windows 10 - WKB119626

Article number: 119626

SYMPTOMS

  • ACQUITY SQD system has software communications issues with trinitiy UI error
  • The tune page becomes unresponsive
  • Issue seen since upgrade to Windows 10
  • Multiple systems affected
  • Seems to be a random occurrence

ENVIRONMENT

  • ACQUITY SQD system
  • Empower
  • Windows 10

CAUSE

User is accessing the acquisition PC via remote desktop with a touch screen monitor.
Remote desktop use is not tested or supported for Empower use (see article WKB50809).

FIX or WORKAROUND

Disable the PC touchscreen as described in article WKB49433.

ADDITIONAL INFORMATION

 