How to restore communication with e2695 and 2998 or 2414 when using Empower - WKB119629

Article number: 119629

OBJECTIVE or GOAL

Restore communication with e2695 and 2998.

ENVIRONMENT

  • Empower 3 Enterprise
  • Empower 3 Personal
  • Alliance e2695
  • 2998 PDA detector
  • 2414 RI detector

PROCEDURE

  1. Power-off the instruments.
  2. Using Empower Configuration Manager, reboot the LAC/E 32 connected to the instruments (or reboot the Empower Personal Workstation).
  3. When the LAC/E 32 has completely booted back up, power-on the instruments.
  4. Allow the instruments to completely initialize.
  5. Open Empower Run Samples and confirm communication.

ADDITIONAL INFORMATION

 