No communication with Acquity modules - WKB107208

Article number: 107208

SYMPTOMS

  • No communication with Acquity modules
  • ICS drivers installed and correct
  • Network card active
  • Exceptions for Antivirus software present
  • No ping responses using NetEvaluator

ENVIRONMENT

  • Empower 3

CAUSE

The Waters DHCP Server configuration wizard was not configured to the Instrument LAN network card

FIX or WORKAROUND

  1. Configure the Waters DHCP Server configuration wizard to the Instrument LAN network card

https://support.waters.com/KB_Inf/Empower_Breeze/WKB7330_Configuring_the_Waters_DHCP_Server

 

ADDITIONAL INFORMATION

https://support.waters.com/KB_Inf/Empower_Breeze/WKB1018_What_is_the_NetEvaluator_utility

 

id107208, comms, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP