What are the ways to optimize peak shape and improve chromatography for nano LC separations? - WKB237325

Article number: 237325

ENVIRONMENT

  • ACQUITY M-Class
  • nanoACQUITY
  • nanoEase M/Z
  • Trap column
  • Method development
  • Peptide separation
  • Chromatography
  • nano LC-MS

ANSWER

Suggestions for optimizing peak shape and improving the chromatography include the following:

  • The age and quality of the mobile phase modifier is important for peptide peak shape and solubility and should be changed out regularly when developing a new method or troubleshooting an established method.
  • If the system setup allows for reverse eluting the Trap column, consider evaluating this option.
  • Ensure that the analytical column is not overloaded; the Trap may have a higher loading capacity than the analytical column doing the separation.
  • Evaluate different column temperatures.
  • Evaluate different column chemistries.
  • Ensure that the analytical flow rate is not too fast for peptide separations.
  • Check gradient slope. Gradients that are too steep will also affect peptide separations.