What are some suggestions for optimizing peak capacity in nano LC separations? - WKB237327

Article number: 237327

ENVIRONMENT

  • nanoEase M/Z
  • ACQUITY M-Class
  • nanoACQUITY
  • Peptide separation
  • Method development
  • Peak capacity
  • nano LC-MS
  • Trap column

ANSWER

Suggestions for optimizing peak capacity include the following:

  • Investigate the use of the CSH C18 chemistry for high peptide peak capacities in the presence of formic acid mobile phases.
  • Investigate the time and flow rate to load the sample on the Trap column if a Trap column is being used. An example of Trap flow loading is 5 to 25 uL/min for 1-3 minutes.
  • Investigate the use of differential column heating on the peptide separation.

Not able to find a solution? Click here to request help.