What are some suggestions for optimizing peak capacity in nano LC separations? - WKB237327
Article number: 237327
ENVIRONMENT
- nanoEase M/Z
- ACQUITY M-Class
- nanoACQUITY
- Peptide separation
- Method development
- Peak capacity
- nano LC-MS
- Trap column
ANSWER
Suggestions for optimizing peak capacity include the following:
- Investigate the use of the CSH C18 chemistry for high peptide peak capacities in the presence of formic acid mobile phases.
- Investigate the time and flow rate to load the sample on the Trap column if a Trap column is being used. An example of Trap flow loading is 5 to 25 uL/min for 1-3 minutes.
- Investigate the use of differential column heating on the peptide separation.
ADDITIONAL INFORMATION
For additional information, see Optimizing Peak Capacity in Nanoscale Trap-Elute Peptide Separations with Differential Column Heating