Adjusting the solvent delivery tube positions

Article number: 81033

For correct operation of the IntelliStart Fluidics system, you must adjust each solvent delivery tube so that its end is close to, but does not touch, the bottom of the reservoir bottle or low volume vial.

To adjust the position of a solvent delivery tube:

  1. Open the access door to the fluidics pump.

SELECT SERIES Cyclic IMS Reservoir Bottle adjustment.png
Callout 1.PNG Access door
Callout 2.PNG Tubing guide
Callout 3.PNGSolvent delivery tube

  1. Move the solvent delivery tube so that its end is close to, but does not touch, the bottom of the reservoir bottle or low volume vial.

  2. Close the access door.