Emptying the nitrogen exhaust trap bottle
Article number: 222864
Inspect the nitrogen exhaust trap bottle in the instrument exhaust line daily, and empty it before it is more than approximately 10% full.
Required materials
- Chemical-resistant, powder-free gloves
To empty the nitrogen exhaust trap bottle:
- Pull the source enclosure release (located at the bottom, right-hand side) outward, and swing open the enclosure.
Warning: To avoid personal contamination with biologically hazardous, toxic, and corrosive materials, wear chemical-resistant, powder-free gloves when performing this procedure.
- Unscrew and remove the nitrogen exhaust trap bottle from the cap and associated fittings.
Warning: To avoid spreading contamination with biologically hazardous, toxic, and corrosive materials, dispose of all waste materials according to local environmental regulations.
- Dispose of the waste liquid in accordance with local environmental regulations.
- Fit and tighten the nitrogen exhaust trap bottle to the cap.
- Secure the nitrogen exhaust trap bottle in the upright position.
- Close the source enclosure.
Tip: An automatic pressure test is performed.