Fitting the ion block support to the source
Article number: 81175
Required materials
- Chemical-resistant, powder-free gloves
- 3-mm hex wrench
- New seals and O-rings
To fit the PEEK ion block support to the source:
Warning: To avoid personal contamination with biologically hazardous, toxic, and corrosive materials, wear chemical-resistant, powder-free gloves when performing this procedure.
- Ensure that the grooves for the PEEK ion block support O-rings are free from dirt and debris.
Tip: If contamination is present, use 1:1 methanol/water, applied to a lint-free cloth, to carefully clean the grooves.
- Fit the O-rings (new ones if you disposed of the old) to the PEEK ion block support.
Tip: To fit an O-ring in its groove, start fitting the O-ring at the notch in the groove, and then progressively work the ring into the groove in either direction from the notch.
- Fit the PEEK ion block support to the instrument’s housing.
- Use the 3-mm hex wrench to fit and tighten the four PEEK ion block support securing screws.
- Fit the ion block assembly to the PEEK ion block support (see Fitting the ion block assembly to the source assembly).
- Fit the source enclosure to the instrument (see Fitting the source enclosure to the instrument).