How to perform a physical reset of the 2767 Sample Manager - WKB97805
Article number: 97805
OBJECTIVE or GOAL
Perform a physical reset of the 2767 Sample Manager.
ENVIRONMENT
- MassLynx
- 2767 Sample Manager
PROCEDURE
- Close MassLynx.
- Power-off the 2767.
- Remove the far-left fraction collection rack.
- Carefully move the Z sample and fraction collection probes up above the height of the fraction collector racks, and then slide the arm to the front-left and down position.
- Carefully move the Z sample and fraction collection probes down to their farthest position.
- Power-on the 2767.
- Open MassLynx and open the inlet editor to reset.