How to perform a physical reset of the 2767 Sample Manager - WKB97805

Article number: 97805

OBJECTIVE or GOAL

Perform a physical reset of the 2767 Sample Manager.

ENVIRONMENT

  • MassLynx
  • 2767 Sample Manager

PROCEDURE

  1. Close MassLynx.
  2. Power-off the 2767.
  3. Remove the far-left fraction collection rack.
  4. Carefully move the Z sample and fraction collection probes up above the height of the fraction collector racks, and then slide the arm to the front-left and down position.
  5. Carefully move the Z sample and fraction collection probes down to their farthest position.
  6. Power-on the 2767.
  7. Open MassLynx and open the inlet editor to reset.

ADDITIONAL INFORMATION

 

id97805, 2767, 2767SFC, MLYNX, MLYNXV41, SUPMM