Pulsing signal with Xevo TQ and ACQUITY UPC2 system - WKB98732

Article number: 98732

SYMPTOMS

  • Pulsing signal with Xevo TQ and ACQUITY UPC2 system
  • ISM pump passes dynamic leak test
  • High pressure (>200 psi) observed from MS capillary when connected directly with ISM

ENVIRONMENT

  • MassLynx
  • Xevo TQ
  • ACQUITY UPC2

CAUSE

The MS capillary is partially blocked.

FIX or WORKAROUND

Replace the capillary tube (700004206).

ADDITIONAL INFORMATION

 

id98732, DLT, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL, XEVOTQ, XEVOTQA, XEVOTQIVD