"Invalid plate position" with 2767 Sample Manager - WKB8489
Article number: 8489
SYMPTOMS
- 2767 invalid plate position error upon initialization
- Plate positions defined in the bed layout are correct
- Reference position could not be calibrated because the icon is unavailable
- Reset of communications and reboot do not remove error
- Most common configuration: 3 x 96 Deep Well Vial Racks and 4 x 18-mm Scintillation Vessel Racks
ENVIRONMENT
- 2767 Sample Manager.
CAUSE
The x axis component for the collection plate is out of range, exceeding the limits of the bed in the software. The limits for the x axis in the software are 0 (min) - 6423 (max).
FIX or WORKAROUND
- Modify the bed layout and set an appropriate number for the X, Y, and Z parameters of each rack. (You may need to manually specify values.)
- Recalibrate the reference position.
- Re-initialize the system.
ADDITIONAL INFORMATION
id8489, 2767, 2767SFC, comms