2525/2545 Binary/Quaternary Gradient Module experiencing high pressure - WKB84293

Article number: 84293

SYMPTOMS

  • 2525, 2535, 2545, 2555, BGM, QGM experiencing higher than normal pressure, even when disconnected from injector valve

ENVIRONMENT

  • 2525/2545 Binary Gradient Module
  • 2535/2545/2555 Quaternary Gradient Module
  • Autopurification
  • FractionLynx
  • Prep 150

CAUSE

T filter is clogged

FIX or WORKAROUND

For 2535/2545/2555 Quaternary Gradient Module:

Order T Filter Assy part 700009202, or Assy, Filter, 19mm, Symmetry, 2 micron (WAT033350)

For 2545 Binary Gradient Module:

Order T Filter Assy part 700001836, or Assy, Filter, 19mm, Symmetry, 2 micron (WAT033350)