"Invalid bed layout" for 2767 - WKB85819
Article number: 85819
SYMPTOMS
- "OK" LED in Inlet Status is red
- Error message reads "2767: Invalid bed layout"
ENVIRONMENT
- Autopurification
- FractionLynx
- 2767 Sample Manager
CAUSE
Collection bed layout x value is outside acceptable limits.
FIX or WORKAROUND
- Open MassLynx software.
- Click on the inlet method.
- Go to bed layout and click on modify bed layout of tray 1 (first collection tray).
- Decrease the x value to 6400.
- Reset injector and recalibrate probe reference position and bed layout.
ADDITIONAL INFORMATION
If x was close to the limit "left" originally, recalibration of probe position can send it out of limit as well.
