No response on any Autopurification system detector - WKB67756
Article number: 67756
SYMPTOMS
- No response on any detector in an Autopurification system
- No response when running prep method, but analytical method works fine
ENVIRONMENT
- Autopurification system
CAUSE
Blocked splitter
FIX or WORKAROUND
Replace the splitter and/or the splitter tubing. (See attached links to Waters Parts Locator for part numbers.)
ADDITIONAL INFORMATION