Error "Unknown Error while processing Raw Data" in FractionLynx - WKB67437
Article number: 67437
SYMPTOMS
-
Error state "Unknown Error while processing Raw Data" present in FractionLynx log
-
Error state occurs consistently during acquisition
ENVIRONMENT
- MassLynx
- FractionLynx
- 2767 Sample Manager
CAUSE
Configuration error in the FractionLynx file (for example, Stop Collect and Start Collect occurring at the same time code)
OR
Corrupted FractionLynx file
FIX or WORKAROUND
Inspect the settings within the FractionLynx file. Either fix the configuration issue OR create a new Fraction file, paying close attention to detail.