Waters Fraction Collector III not collecting fractions - WKB231286

Article number: 231286

SYMPTOMS

  • System was recently relocated
  • Waters Fraction Collector III is not collecting fractions, needs to be started manually

ENVIRONMENT

  • Waters Fraction Collector III
  • Empower 3

CAUSE

The EXT.START cables are connected incorrectly.

FIX or WORKAROUND

Confirm that EXT.START cable is connected as follows: black connected to 3(-), red connected 4(+)

ADDITIONAL INFORMATION