2767 Sample Manager unable to aspirate sample - WKB232860
Article number: 232860
SYMPTOMS
- 2767 Sample Manager unable to aspirate sample
ENVIRONMENT
- AutoPurification
- FractionLynx
- 2767 Sample Manager
CAUSE
The T7 tubing connected to the sample needle is blocked.
FIX or WORKAROUND
- Remove the sample needle as described in the 2767 Owner Manual.
- Power-on the 2767 and flush the T7 tubing by going to Inlet Method Editor > Waters2767 and selecting Probe/Port Wash and Wash Right Injector.
- Hold the T7 tubing in the wash station to allow the sample to be dislodged.
- Perform Needle Calibration as described in the 2767 Owner Manual.
ADDITIONAL INFORMATION

