Shifts in retention times on Fluid Delivery Module 10 - WKB226946
Article number: 226946
SYMPTOMS
- Shifts in retention times on Fluid Delivery Module 10
- Some peaks elute earlier, some peaks elute later and are reproducible
- The CO2 inlet temperature readback fluctuates between 12.0 and13.5 °C
- The Pressure Drop readback fluctuates between 0 and 10 bar
ENVIRONMENT
- FDM
- MassLynx
CAUSE
Failing check valve
FIX or WORKAROUND
Replace the check valve or valves (part 700008840).
ADDITIONAL INFORMATION
The CO2 inlet temperature readback should be approximately 11 °C with flow on.
The pressure drop delta (at constant conditions) should be approximately 1-2 bar depending on co-solvent composition.