Air bubbles in syringe of 2767 sample manager - WKB229286
Article number: 229286
SYMPTOMS
- Air bubbles present in sample syringe and tubes
- Multiple system primes do not correct issue
ENVIRONMENT
- 2767 Sample Manager
CAUSE
The nuts on the syringe valve and wash pump fitting are loose.
FIX or WORKAROUND
1. Tighten the nuts on the syringe valve and wash pump.
2. Prime the system until the syringe and tubes are filled with fluid.
ADDITIONAL INFORMATION
id229286, 2767, 2767SFC