Skip to main content
Waters

Air bubbles in syringe of 2767 sample manager - WKB229286

Article number: 229286

SYMPTOMS

  • Air bubbles present in sample syringe and tubes
  • Multiple system primes do not correct issue

ENVIRONMENT

  • 2767 Sample Manager

CAUSE

The nuts on the syringe valve and wash pump fitting are loose.

FIX or WORKAROUND

1. Tighten the nuts on the syringe valve and wash pump.

2. Prime the system until the syringe and tubes are filled with fluid.

clipboard_e51e503375428bb19331a539e0fa85356.png

 

ADDITIONAL INFORMATION

 

id229286, 2767, 2767SFC

Not able to find a solution? Click here to request help.