Unable to view or edit methods in ChromScope 2.1 - WKB225748
Article number: 225748
SYMPTOMS
-
Unable to load, view, or edit one or more instrument, processing, or collection methods in any project
-
Method tabs missing or blank
-
Gives a ChromScope Exception of "This operation is invalid since the DockingWindow has not yet been registered with a DockSite. (ChromScope.IU)"
-

ENVIRONMENT
- ChromScope 2.1
- Windows 10
- Investigator SFC
CAUSE
Windows updates or other security software settings may have corrupted or are blocking ChromScope functions.
FIX or WORKAROUND
- Verify that antivirus/malware and other security software is not installed, or that it has the correct exceptions configured for ChromScope.
- Reinstall ChromScope.
ADDITIONAL INFORMATION
Antivirus exceptions are "c:\Program Files (x86)\ChromScope IE" and "c:\Program Files (x86)\Waters Instruments".
Reinstallation of RocketPort driver and serial ports configuration is usually not necessary.