2767 Sample Manager leaks from wash pass valve - WKB93575
Article number: 93575
SYMPTOMS
- Leakage from wash pass valve
- Headspace forms in the syringe
ENVIRONMENT
- 2767 Sample Manager
- 2525/2545 Binary Gradient Module
- ACQUITY UPLC ELS Detector
- 3100 Mass Detector
CAUSE
Clogged sample probe—the wash pass valve malfunctioned due to high pressure when expelling liquid from the syringe.
FIX or WORKAROUND
- Replace the wash pass valve (700002120).
- Select Waters 2767 > Prime System in Inlet Editor and confirm that there is no leakage in the flow path and no head space in the syringe.
ADDITIONAL INFORMATION
https://support.waters.com/KB_Inst/Mass_Spectrometry/WKB88971_2767_not_injecting_the_correct_volume
https://support.waters.com/KB_Inst/Purification_SFx/WKB14782_2767_Syringe_Command_Overflow