Skip to main content
Waters

2767 syringe: Command Overflow - WKB14782

Article number: 14782

SYMPTOMS

2767 syringe: Command Overflow

ENVIRONMENT

  • 2767 Sample Manager

CAUSE

Syringe motor failure; inlet is blocked

FIX or WORKAROUND

  1. Determine if the path from the 2767 syringe outlet to the syringe needle is blocked (note that there may be sample precipitation).
  2. Determine if the inlet is blocked or damaged.
  3. Replace the syringe pump assembly (700003531).

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.