2767 not injecting the correct volume - WKB88971
Article number: 88971
SYMPTOMS
2767 not aspirating the correct volume.
ENVIRONMENT
- 2767 Sampler Manager
- MassLynx
CAUSE
2767 vented probe was blocked
FIX or WORKAROUND
- Replace Vented Probe Part 700001387
ADDITIONAL INFORMATION
Note: Inspect T7 for blockages
id88971, 2767, 2767SFC, FLYNXV41, MLYNX, MLYNXV41, SUPMM, UNIFSW17