2767 not injecting the correct volume - WKB88971

Article number: 88971

SYMPTOMS

2767 not aspirating the correct volume.

ENVIRONMENT

  • 2767 Sampler Manager
  • MassLynx

CAUSE

2767 vented probe was blocked

FIX or WORKAROUND

  1. Replace Vented Probe Part 700001387

ADDITIONAL INFORMATION

Note: Inspect T7 for blockages 