MassLynx states "acquiring" but does not inject - WKB88943
Article number: 88943
SYMPTOMS
- Inlet method status is ready
- Communications to all modules
- Inlet method can be loaded
- When running a sample, MassLynx states "acquiring" but the bottom status bar remains at "instrument present"
ENVIRONMENT
- ACQUITY QDa Mass Detector
- MassLynx 4.2
CAUSE
The ACQUITY QDa Mass Detector is in an unknown error state.
FIX or WORKAROUND
- Power-cycle the QDa to clear the error state.
- Reboot the PC.
ADDITIONAL INFORMATION
id88943, comms, MLYNX, MLYNXV41, QDA, SUPMM