Skip to main content
Waters

Noisy baseline and low sample energy on TUV - WKB296876

Article number: 296876

SYMPTOMS

  • Noisy baseline
  • Sample energy low, but reference is high

ENVIRONMENT

  • 2489
  • ACQUITY TUV 

CAUSE

Dirty flow cell 

FIX or WORKAROUND

  • Flush the flow cell with 70:20 H20:Phosphoric acid using 100uL syringe, repeat this 5 times over 20mins.
  • Now do the sample but use IPA.
  • Finally flush flowcell with 100% H20.
  • Sample erergy and baseline now much improved.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.