ACQUITY UPLC Column Heater/Cooler (the 30-cm CHC) does not appear in console - WKB296087
Article number: 296087
SYMPTOMS
- Column heater/cooler does not appear in ACQUITY console
ENVIRONMENT
-
ACQUITY Arc Sample Manager FTN-R
-
ACQUITY UPLC 30-cm Column Heater/Cooler (30-cm CHC)
- Alliance Series Column Heater/Cooler
CAUSE
Instruments powered-up in the wrong order
FIX or WORKAROUND
- Power-off the 30-cm column heater/cooler on the side.
- Power-off all ACQUITY modules.
- Power-on the 30-cm column manager on the side.
- Power-on the sample manger first, and then all other ACQUITY modules.
- Confirm that column heater readback is visible in ACQUITY console (in Sample Manager).