How to replace the flow cell on a TUV/PDA - WKB295516

Article number: 295516

OBJECTIVE or GOAL

Replace the flow cell on a TUV/PDA.

ENVIRONMENT

  • ACQUITY Premier PDA
  • ACQUITY Premier TUV
  • ACQUITY TUV
  • ACQUITY PDA

PROCEDURE

  1. Turn off the lamp.
    flowcell.png
  2. Remove the connection with the green mark. 
  3. Remove the tubing from inlet and outlet (blue circles).
  4. Loosen the three screws (red marking). They cannot be removed. 
  5. Replace the flow cell.
  6. Flush the flow cell with organic (e.g., ACN) for a minimum of 15 minutes.
  7. Power-on the lamp, let it warm up, and then calibrate. 