Skip to main content
Waters

Run Samples blank in Empower for Agilent systems - WKB206530

Article number: 206530

SYMPTOMS

  • When connecting Agilent systems to Empower, Run Samples is blank, showing only gray panels
  • DCOM settings correct per WKB121320 
  • .NET Framework settings correct per WKB48362

ENVIRONMENT

  • Windows 10
  • Agilent ICF
  • Empower 3

CAUSE

Drivers were incorrectly installed and not fully deployed.

FIX or WORKAROUND

  1. Move drivers from shared drive to local desktop and install ICF.
  2. Check add/remove programs to ensure that all drivers have deployed after installation.

ADDITIONAL INFORMATION

WKB121320 - How to set DCOM and Local security policy settings for Agilent LC/GC, ICF-based instruments

WKB13765 - How to determine which .NET Framework versions are installed

WKB48362 - What version of Microsoft .NET Framework is required for ICF Support?

id206530, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP

Not able to find a solution? Click here to request help.