Skip to main content
Waters

SM-FTN System leak - WKB199280

Article number: 199280

SYMPTOMS

  • System leaks from the bottom right-hand side of the sample manager

ENVIRONMENT

  •  SM-FTN

CAUSE

The SM-FTN drip tray curtain is not fitted onto the bottom of the lower drip tray. Heater/chiller condensate is seen "leaking".

FIX or WORKAROUND

Fit the SM-FTN drip tray curtain (415001552).

ADDITIONAL INFORMATION

clipboard_ea0d313020532c54d50e9d4b09d4612a1.png

Not able to find a solution? Click here to request help.