How to reset the 2777 plunger stroke limit - WKB19914

Article number: 19914

OBJECTIVE or GOAL

Change the plunger stroke limit to a new value when you see a "plunger stroke limit" message on the 2777 hand-held.

ENVIRONMENT

  • 2777
  • HDX
  • Open Architecture

PROCEDURE

  1. On the hand-held, go to extended user mode.
  2. Press F1 for “Menu”.  
  3. Scroll down to “Info”.
  4. Scroll down to ”Maintenance”.
  5. Reset the 2777 plunger stroke limit to a new value.

ADDITIONAL INFORMATION

  • This is a warning message and the system will run fine.
  • It is simply a counter that can be set in the firmware to give a reminder to change the plunger.