Errors "Low desolvation gas flow" and "The desolvation gas flow is too low" with Xevo TQD - WKB99395
Article number: 99395
SYMPTOMS
- Error" The desolvation gas flow is too low"
- Error "Low desolvation gas flow"
- Error "Stop flow input activated" on solvent manager
- Nitrogen inlet pressure OK
- No leaks seen on system
- The desolvation gas flow parameter in the MS tune page is unresponsive and stuck at a constant value, regardless of what value is entered
ENVIRONMENT
- MassLynx
- Xevo TQD
CAUSE
The Xevo TQD has partially lost communication.
FIX or WORKAROUND
Perform an electronic reset on the MS detector:
- Close MassLynx.
- With a piece of PEEK tubing, push the RESET button on the front of the MS.
- Wait at least three minutes for the MS to initialize.
- Open MassLynx.
- Open the MS Tune page and confirm that the desolvation gas flow readback changes with a newly entered value.
ADDITIONAL INFORMATION
id99395, comms, MLYNX, MLYNXV41, SUPMM, XEVOTQD, XEVOTQDIVD