Source pressure leak warning - WKB99422

Article number: 99422

SYMPTOMS

  • Source pressure leak warning appears during source pressure test.
  • Source high pressure warning.
  • May hear leak noise at the source enclosure.
  • Engineer dashboard looks like waste bottle error.

ENVIRONMENT

  • Xevo QTof
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • SYNAPT XS
  • Vion IMS QTof
  • Universal Lockspray source

CAUSE

Leaking at the Universal Lockspray source O-ring, either from wear of the O-ring or more likely the plastic casing.

FIX or WORKAROUND

Replace the O-ring (5711032-S).

Replace the Lockspray union cover holder (need to order full door casing, not available on own).LockSpray Source Covers (# 700004702)

ADDITIONAL INFORMATION

clipboard_e73ad04f7a217caa225b5f7a19b91a068.png

Engineer dashboard showing leak from Lockspray fitting.

Not able to find a solution? Click here to request help.