"Source pressure test failed - Source pressure test exhaust trip" errors - WKB97941

Article number: 97941

SYMPTOMS

  • "Source pressure test failed - Source pressure test exhaust trip" message
  • "Exhaust pressure too high" warning message
  • "Exhaust Problem Warning" warning message

ENVIRONMENT

  • ACQUITY QDa
  • Xevo G2
  • Xevo G2 S
  • Xevo G2 xs
  • Synapts si
  • Xevo TQ-S Micro
  • Xevo TQ-S
  • Xevo TQ-XS
  • Xevo TQD
  • Xevo TQ Absolute
  • ACQUITY RDa Detector

CAUSE

Faulty trap bottle

FIX or WORKAROUND

  1. Replace the trap bottle assembly.

ADDITIONAL INFORMATION

Part number for QDa trap bottle (700009635)

Part number for RDa trap bottle (700012115)

Part number for Xevo trap bottle (700004606)

Part Number for the Xexo TQD trap bottle (700005811)

Part Number for the Xexo TQ Absolute trap bottle (700013610)

id97941, QDA, RDA, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, XEVOG2XSQT, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD