Ebara backing pump unable to start after power outage - WKB97862
Article number: 97862
SYMPTOMS
-
Unable to pump down instrument after power outage
-
After pressing Pump, turbomolecular pumps start to speed up, but the Ebara pump seems stalled
-
Error code can be seen on Ebara display (Err)
ENVIRONMENT
-
Ebara
-
Vion
-
Xevo TQ-XS
- Xevo TQ-S
CAUSE
The Ebara pump is in an error state as a result of power outage.
FIX or WORKAROUND
- Switch the Ebara pump off.
- Wait for five minutes.
- Switch the Ebara pump on. Check the display on the Ebara pump; you should see "n5500".
-
Pump down the instrument as usual.
ADDITIONAL INFORMATION
It is not necessary to reboot MS instrument electronics during an Ebara pump restart.
id97862, EBARA-SA20, EBARA-SA30, turbo pump, VION, XEVOG2XSQT, XEVOTQS, XEVOTQSIVD, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD