IntelliStart error "Error creating file ParentOptimizeES+ Creation failed" or "Error creating data file AutotuneSourceES+ Creation failed" - WKB9739

Article number: 9739

SYMPTOMS

  • When running IntelliStart "sample tune and develop method", the error message "Error creating file ParentOptimizeES+ Creation failed" or "Error creating data file AutotuneSourceES+ Creation failed" appears

ENVIRONMENT

  • Xevo TQ-S micro
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-MS
  • ACQUITY Console
  • IntelliStart

CAUSE

The data folder is missing from Intellistart.pro.

FIX or WORKAROUND

Manually create a folder called "Data" within the IntelliStart project (Intellistart.pro).