QDa fails positive mass accuracy test; mass accuracy is more than 200 mDa - WKB97401
Article number: 97401
SYMPTOMS
- Mass accuracy test over five compounds (acetaminophen, caffeine, sulfadimethoxine, reserpine, terfenadine) always fails
- Some or all measured masses > 200 mDa compared with theoretical
ENVIRONMENT
- ACQUITY QDa Mass Detector
- Empower 3
- Qualification tests
- Startup tests
CAUSE
Sampling rate in the instrument method is higher than 5 points/sec.
Example: At 10 points/sec, mass accuracy test fails and most of the measured masses are over 200 mDa specification.
FIX or WORKAROUND
Set 5 points/sec in the sampling rate feature to be accurate against the specifications and to pass the test.
ADDITIONAL INFORMATION
id97401, EMP2QLIC, EMP3Q1ULIC, EMP3QLIC, QDA, UNIFQLIC